There are BJH, CI, and DH method (for cylindrical pore shape) and Innes method (for slit pore shape) as the mesopore analysis. All these theories are based on the calculation of capillary condensation theory (Kelvin equation). Generally, this theory is applied to the pores above 2 nm.
The pore shape model needs to be taken into the account, which is the same for micropore analysis. BJH, CI, and DH method will obtain mostly the same result. The only difference of these theories is the correction term of adsorption layer thickness ft(P / P0). This difference becomes prominent below 2 nm (non-applicable range). These theories are also applicable to macropores, but there is an upper limitation due to the measurement capability. As described, the pore size is calculated from Kelvin equation. In case of the pore size over 100 nm, it is difficult to measure the relative pressure of around P/P0=0.99 with good accuracy. The bigger the pore size, the more difficult it becomes to achieve the enough accuracy of the measurement.